Resumo do Conteúdo

Página 1 - Paradigm™ Infusion Pump

Paradigm™ Infusion PumpModel MMT-511User Guide

Página 2

vi Adverse reactions. . . . . . . . . . . . . . . . . . . . . . . . . . . . 130Hyperglycemia and diabetic ketoacidosis (DKA) . . 130Hypoglycemia

Página 3 - Electrostatic Discharge

88 Chapter 7 BlockThe block feature restricts access to the pump programming. This is an important safety feature if the pump users are young children

Página 4

Utilities 89 2. Press the button to select Block. Press ACT.3. Press the button to select ON. Press ACT.When the Block Option is ON, the pump wi

Página 5 - Table of Contents

90 Chapter 7 Setting the time and date1. From the HOME screen, press ACT. Press the button to select Utilities. Press ACT.2. Press the button to

Página 6

Utilities 91 5. The hour will flash. Press the and buttons to select the hour. Press ACT to set the hour.When pressing the and buttons the AM

Página 7 - Chapter 6

92 Chapter 7 8. The month will flash. Press the and buttons to select the month. Press ACT.9. The day will flash. Use the and buttons to selec

Página 8 - Chapter 8

Utilities 93 LanguageThe language shown on the screens on the pump can be changed. Some languages will not be available on all pump models. Setting th

Página 9 - Chapter 10

94 Chapter 7 RF options (remote control)➤What are RF Options?RF Options allow you to program your pump by using a remote control(s).CAUTION: The remot

Página 10 - Chapter 11

Utilities 95 2. Press the button to select RF Options. Press ACT.3. Press the button to select ON. Press ACT.4. ADD ID will be selected. Press

Página 11 - Chapter 12

96 Chapter 7 6. The RF ID is printed on the back of the remote control.Deleting an RF IDEnables you to delete an RF ID code you have set.1. From the

Página 12

Utilities 97 3. ON will be selected. Press ACT.4. Press the button to select Delete ID. Press ACT.5. Select the RF ID code you want to delete. Pre

Página 13 - Introduction

vii Chapter 12Pump Specifications . . . . . . . . . . . . . . . . . . . . 143Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 14 - Availability

98 Chapter 7 Reviewing the RF ID codesAllows you to review the RF ID codes you have already programmed.1. From the HOME screen, press ACT. Press the

Página 15 - Emergency kit

Utilities 99 4. Press the button to select Review ID. Press ACT.5. Review the RF ID codes you have set.Clear pumpIn rare instances, you may want t

Página 16 - 4 Chapter 1

100 Chapter 7 Clearing pumpClears the programming you have set and restores the factory default settings stored in the pump memory, except for langua

Página 17 - Your pump

Utilities 101 4. The pump will ask you to Confirm that you want to Clear the pump. Select the option you desire.5. Press ACT. A message will appear

Página 18 - 6 Chapter 1

102 Chapter 7 8. Disconnect pump from your body. Remove reservoir. Press ACT to rewind.9. The pump will now rewind.10. When the pump has finished re

Página 19 - Introduction 7

Utilities 103 12. Select YES and press ACT to verify insulin type. To change the insulin type, select NO and press ACT, or press ESC to return the IN

Página 20 - Modes of operation

104 Chapter 7 If the pump detects a condition; such as, low battery, the test will not finish. A message will appear indicating the condition that cau

Página 21 - Attention

Utilities 105 3. Screen Test: The screen should appear as shown below (all black).4. Selftest: The pump will count down from 10.5. Tone Test: You s

Página 22 - Pump buttons

106 Chapter 7 7. Once the selftest is completed, TEST OK will appear on the screen.8. Contact the 24-hour Product Help-line if any of the tests do n

Página 23 - Introduction 11

107 Chapter 8• • • • • • Remote ControlCAUTION: The remote control should not be used aboard an air-craft. Operate your pump using the buttons on

Página 25 - Suspend

108 Chapter 8 Using your remote controlDiscuss what uses and settings are best for you with your health-care professional. See Chapter 7, Utilities fo

Página 26 - Utilities menu

Remote Control 109 Waking up your pumpThe remote control will operate within 24 inches from your pump. When the remote control is in “sleep mode,” the

Página 27 - Initial pump settings

110 Chapter 8 Deliver an easy bolus1. With pump in the HOME screen, press and hold ACT to “wake up” the pump.2. When the pump is awake, it will beep

Página 28 - Blood glucose targets

Remote Control 111 3. To resume basal delivery, press and hold ACT to “wake up” the pump. Press the S button, then press ACT. The pump will beep/vibr

Página 29 - Getting Started

112 Chapter 8 2. Make sure the RF ID is correctly entered.3. Make sure the pump is not in a “Low Batt” condition. 4. Call the 24-hour Product Help-

Página 30 - Preparing your pump for use

113 Chapter 9• • • • • • Troubleshooting & CareIt is recommended that you read your warranty statement included with your pump for information

Página 31

114 Chapter 9 5. Continue troubleshooting by disconnecting at the quick-disconnect and set a 10 unit Fixed Prime. 6. Does insulin come out of the n

Página 32

Troubleshooting & Care 115 If the pump goes to the REWIND screen after you clear the alarm, you will need to perform a rewind per the instructions

Página 33 - Installing the reservoir

116 Chapter 9 6. Perform a selftest.My screen appears distortedThe screen may appear distorted or have a “rainbow” appearance if you are wearing pola

Página 34 - Rewinding the pump

Troubleshooting & Care 117 Pump maintenanceBatteryCertain features on the Paradigm pump use a lot of battery power. Remote control Backlight Vibr

Página 35 - Verifying insulin type

1 Chapter 1• • • • • • IntroductionWhether you are beginning pump therapy for the first time, or upgrading from a previous model, we are pleased t

Página 36 - Changing insulin type

118 Chapter 9 out for more than one (1) minute or if it was placed back-wards in the pump.1. Using the edge of a coin, unscrew the battery cap in a c

Página 37 - Hold ACT to Prime

Troubleshooting & Care 119 Cleaning your pump1. Use only a damp cloth and mild detergent mixed with water to clean the outside of your pump. 2.

Página 38 - MANUAL PRIME

120 Chapter 9 PrecautionsAvoiding extreme temperatures1. Avoid exposure of your pump and remote control to temperatures above 104°F (40°C) or below

Página 39 - Getting Started 27

Troubleshooting & Care 121 5. Perform the Selftest procedure located in the Utilities menu. 6. Call the 24-hour Product Help-line for assistance

Página 41

123 Chapter 10• • • • • • User Safety IndicationsThe Paradigm Insulin pump is indicated for the continuous delivery of insulin, at set and variab

Página 42 - Priming with fixed prime

124 Chapter 10 WarningsDiabetic ketoacidosis (DKA)Insulin pump therapy uses only regular (short-acting) insulin; therefore, any interruption in the de

Página 43 - PRIME DELIVERY

User Safety 125 Reservoir and infusion setsStandard Luer infusion sets are not compatible with the Medtronic MiniMed Paradigm pump. Medtronic Minimed

Página 44 - Reviewing the prime history

126 Chapter 10 ✒Act quickly to respond to abnormal blood glucose. Notify your healthcare professional of low blood glucose requiring assistance or of

Página 45 - How to wear the pump

User Safety 127 Precautions - hypoglycemiaTo help prevent hypoglycemia (low blood glucose), routinely follow these precautionary steps:1. Know the s

Página 46 - 34 Chapter 2

2 Chapter 1 AssistanceMedtronic MiniMed provides a 24-hour help line for assistance in the United States. The help line is staffed with technicians wh

Página 47 - Status

128 Chapter 10 Precautions - hyperglycemiaTo help prevent hyperglycemia (high blood glucose), routinely fol-low these precautionary steps:1. Know the

Página 48 - 36 Chapter 3

User Safety 129 Precautions - infusion sets and sitesUse only Paradigm style reservoirs labeled “for use with MiniMed Paradigm pumps.” Do not use oth

Página 49 - Bolus Menu

130 Chapter 10 Adverse reactionsHyperglycemia and diabetic ketoacidosis (DKA)Pump wearers use only regular short-acting insulin. As a result, they do

Página 50 - SET BOLUS

User Safety 131 HypoglycemiaHypoglycemia can be more subtle for pump users than with patients on conventional therapy. It may occur even when the pump

Página 51 - Delivering a normal bolus

132 Chapter 10 NoticeInsulin pump and remote controlThese devices comply with the United States Federal Communica-tions Commission and international s

Página 52 - Dual/Square bolus

User Safety 133 CAUTION: Any changes or modifications to the devices not expressly approved by Medtronic MiniMed could void your ability to operate th

Página 53

134 Chapter 10

Página 54 - 42 Chapter 4

135 Chapter 11• • • • • • Alarms and AlertsYour Paradigm pump has a sophisticated network of safety checks and systems.If the safety network detec

Página 55 - Bolus Menu 43

136 Chapter 11 When you see a alarm message on your pump’s screen:1. Use the button to scroll down to read ALL of the alarm text. There is often tex

Página 56

Alarms and Alerts 137 Auto off This alarm alerts you to the fact that no buttons have been pressed during the time limit that you set in the AUTO OFF

Página 57 - Bolus Menu 45

Introduction 3 Emergency kitKeep an emergency kit with you at all times to make sure that you always have necessary supplies. This kit should include:

Página 58 - Dual wave bolus

138 Chapter 11 No delivery Insulin delivery has stopped. This alarm will occur if your pump detects a blockage. Motor errorInsulin delivery has stoppe

Página 59 - Dual Now

Alarms and Alerts 139 Max delivery This alarm alerts you when you have taken more insulin than expected based on maximum bolus and maximum basal rates

Página 60

140 Chapter 11 Bolus stoppedIf this alarm should occur, it is very important to check bolus history to review how much of the bolus was actually deliv

Página 61 - Delivering an express bolus

Alarms and Alerts 141 AlertsSilence the Alerts by pressing ESC, ACT to clear. Follow the instructions on the screen. Low battery After receiving this

Página 62 - Dual Wave Bolus

142 Chapter 11 Message in Alarm HistoryAfter putting a battery into your new pump, a message “RfBad” may appear in your alarm history screen. (The pum

Página 63 - Bolus Menu 51

143 Chapter 12• • • • • • Pump Specifications The Pump Specifications section provides detailed information on specifications related to your pump

Página 64

144 Chapter 12 Insulin delivery Allows up to 48 individual basal rates to be programmed in a daily pattern. Different basal rates can be programmed ev

Página 65 - Setting up easy bolus

Pump Specifications 145 Easy bolus Features programming using audible tones (or vibrate pulses) in user-determined increments. Beep mode range: 0 to

Página 66 - 54 Chapter 4

146 Chapter 12 Occlusion detectionThe detection of an occlusion will trigger a “no delivery alarm.” Up to 3.0 units of “missed” insulin are required t

Página 67 - Delivering an easy bolus

Pump Specifications 147 Prime function Is used to fill the infusion set tubing prior to use. Primes are not counted in daily totals but are recorded s

Página 68 - Bolus history

4 Chapter 1 Pump and user guide conventionsThe following terms and icons are used to describe the pump in this User Guide. “Press” means to push and r

Página 69

148 Chapter 12 Square wave bolus Delivers 0.1-25.0 units of insulin over a duration of 30 minutes to eight hours (limited by maximum bolus setting).Te

Página 70 - 58 Chapter 4

Pump Specifications 149 Factory default settingsThe following are the default settings:Menu DefaultBolus Menu: Easy Bolus: ON Increment default: 0.5 u

Página 71 - Basal

150 Chapter 12 Icon tableDo not reuse:Attention: See Instructions for UseMethod of sterilization using Ethylene Oxide:Date of Manufacture: (year - mo

Página 72

Pump Specifications 151 Language of Instructions for UseRecycleKeep dryF

Página 74 - Delivering a basal

153 • • • • • • IndexAadverse reactions 130Alarms 81, 135, 143auto-off 137batt out limit 138button error 139check settings 139E alarm 136history 86

Página 75 - BASAL RATE

start and stop times 61basal 61basal menu 14Batt 138Batterypump 114, 117, 146remote control 109, 111blockwhen available 88block feature 144blood glu

Página 76 - Setting max basal rate

delivery accuracy 144diabetic ketoacidosis 124, 130displacement accuracy 144DKA 124drive motor 144dropped 120dual wave bolus 37, 46, 144Eeasy bolus 3

Página 77 - BASAL MENU

insulin 129, 130, 145insulin delivery 144Kketones 130Llanguage 93Mmaximum bolus limit 37, 50Modesattention 9normal 8special 8MRI 125Nno delivery 113

Página 78 - How does it work?

priming 21pump buttons 10pump status 35Rremote control 94, 107reservoir 5, 129, 147rewind 21, 22RF ID 94Ssafety 147safety checks 147selftest 103, 105

Página 79

Introduction 5 Your pumpTake a look at your pump. The pump is smart. It remembers your settings and stores a record of its actions.Turn your pump on i

Página 80 - Basal 1: 0.8 U/H

time/date 18, 90Uutilities 81Wwarnings 124water 112watertight 148Xx-rays 125

Página 81 - Personal basal patterns

6 Chapter 1 Installing the battery the first timeThe Paradigm pump uses a AAA alkaline battery. As a safety mea-sure, Medtronic MiniMed has designed t

Página 82 - 70 Chapter 5

Introduction 7 3. Numbers will appear on the pump screen as the pump turns on. Then the following screen will appear. 4. If this does not happen: a.

Página 83

© 2005, Medtronic MiniMed. All rights reserved.This product is covered by U.S. Patent Nos. 6,551,276 and 6,554,798. Other U.S. and/or foreign paten

Página 84

8 Chapter 1 Home screenThe HOME screen serves as the start-ing point to access the programming screens by pressing ACT, or to access the STATUS screen

Página 85 - Select Patterns

Introduction 9 The following features or alarms turn on the open circles: Dual/square wave bolus delivery Basal patterns A or B Temporary basal delive

Página 86 - Basal review

10 Chapter 1 The screens that will display solid circles are:Pump buttonsFive buttons are used to program and access the pump features. The UP and DO

Página 87 - Basal 75

Introduction 11 From the HOME screen, the and buttons have special functions:✒ The button lets you set an Easy Bolus.✒ The button lets you tu

Página 88 - 76 Chapter 5

12 Chapter 1 Main menuWhen you press ACT from the HOME screen, the Main menu will appear. MiniMedMAIN MENUBolusSuspendBasalPrimeUtilitiesMiniMedBOLUS

Página 89 - Press ACT to Stop Pump

Introduction 13 Five lines of text is always viewable on your pump screen. The top line will always display the word “MiniMed.” The lines that follow

Página 90 - Suspending the pump

14 Chapter 1 Basal menuUse the Basal menu to program a continuous amount of insulin 24 hours-a-day. Your basal rate is the amount of insulin required

Página 91 - Resuming pump delivery

Introduction 15 Initial pump settingsThis is the information you need on the day that you begin using your pump with insulin. If your pump start is to

Página 92 - 80 Chapter 6

16 Chapter 1 Basal rate informationMost people will only have one basal rate when starting pump therapy. Meal bolusesCorrection bolusBlood glucose ta

Página 93 - Utilities

17 Chapter 2• • • • • • Getting StartedThis chapter will guide you through the set-up procedures to get your pump ready to use. Make sure you have

Página 94 - Setting the alert type

Important Safety Information Regarding Your Paradigm Insulin Infusion Pump (includes all models)Avoid Immersing Your Pump In WaterAlthough it is unlik

Página 95 - Press ACT

18 Chapter 2 Refer to the Instructions for Use included with the Paradigm reser-voir and infusion sets you have chosen for detailed information.Prepar

Página 96 - Using Auto-off

Getting Started 19 3. Press the and button to select your preferred time format. Press ACT.If you have selected the 12 Hour Setup, when pressing

Página 97 - Reviewing the alarms

20 Chapter 2 6. The minutes will flash. Press the and buttons to set the minutes. Press ACT.7. The year will flash. Press the and buttons to

Página 98 - Daily totals

Getting Started 21 10. The time and date will be shown.11. Press the ACT button to return to the UTILITIES menu.Installing the reservoirIf using the p

Página 99 - Reviewing the daily totals

22 Chapter 2 Removing the reservoir1. Remove the infusion set from your body or disconnect the tubing from the infusion site.2. Using a counter-clo

Página 100 - Setting up block

Getting Started 23 WARNING: Never rewind the pump while it is connected to your body or contains a reservoir. Never insert the reservoir into pump whi

Página 101 - BLOCK OPTION

24 Chapter 2 7. Use the and buttons to select your reservoir type/insu-lin concentration. Press ACT.8. Use the button to select Yes and press

Página 102 - Setting the time and date

Getting Started 25 10. Press ESC to return to the INSULIN TYPE screen to change your Insulin Type if you have selected the wrong insulin type. 11. Pre

Página 103 - 10:12 PM

26 Chapter 2 16. Fill the reservoir with your insulin. Refer to the Instructions for Use provided with the reservoir.WARNING: Your pump will not accu

Página 104 - 92 Chapter 7

Getting Started 27 WARNING: Improper placement of the reservoir may result in inac-curate delivery of insulin. Check to make sure the tubing connector

Página 106 - Adding an RF ID

28 Chapter 2 Manual Prime1. Make sure that the infusion set is disconnected from your pump to your body.2. Hold ACT to Prime. The pump will prepare

Página 107 - ADD RF ID

Getting Started 29 4. After the second set of six beeps, the number of units used will count up on the screen and begin filling the tubing with insu-

Página 108 - Deleting an RF ID

30 Chapter 2 Priming with fixed primeFixed priming provides a pre-determined volume for priming the infusion set soft cannula after the introducer nee

Página 109 - DELETE RF ID

Getting Started 31 Manual Prime will ONLY appear in the Prime menu after you have performed “Rewind” when changing your reservoir. The Manual Prime fe

Página 110 - Reviewing the RF ID codes

32 Chapter 2 If you need to stop a Fixed Prime, select Suspend, and then Resume. Reviewing the prime historyReview the prime history to see the primes

Página 111 - Clear pump

Getting Started 33 3. Use the and buttons to review the primes. The “F” at the end of the line indicates a fixed prime. An “M” indicates a manual

Página 113

35 Chapter 3• • • • • • Status Checking pump statusThe different operations of the Paradigm pump can be easily reviewed by accessing the STATUS sc

Página 114 - INSULIN TYPE

36 Chapter 3 The following list is information that appears on the STATUS screen. All this information will not appear all the time or at the same tim

Página 115 - Selftest

37 Chapter 4• • • • • • Bolus MenuDuring a bolus delivery most pump features are disabled until after the bolus has been delivered. Suspend and th

Página 116 - MAIN MENU

i • • • • • • Table of ContentsChapter 1Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . 1For your records . . . . . . . . .

Página 117 - Utilities 105

38 Chapter 4 * The SET BOLUS screen is the default Bolus screen. If Dual/Square is ON the BOLUS TYPE MENU screen will appear.MiniMedBOLUS MENUBolus H

Página 118 - 106 Chapter 7

Bolus Menu 39 Delivering a normal bolus1. From the HOME screen, press ACT. Bolus will be selected. Press ACT. 2. The Set Bolus option will be select

Página 119 - Remote Control

40 Chapter 4 4. The bolus will be delivered now. A beep or vibration will occur at the start and end of the bolus. As the bolus is delivered, the amo

Página 120 - Using your remote control

Bolus Menu 41 3. Select ON. Press ACT. 4. Repeat these steps to turn OFF this feature if desired.Square wave bolusThe square wave bolus can be used f

Página 121 - Waking up your pump

42 Chapter 4 Delivering a square wave bolusA square wave bolus delivers a bolus evenly over a period of time, from 30 minutes to 8 hours.During a Squa

Página 122 - Deliver an easy bolus

Bolus Menu 43 4. The SET SQUARE BOLUS screen will appear. The bolus amount will flash. Use the and buttons to set the bolus. Press ACT.5. The du

Página 123 - Remote Control 111

44 Chapter 4 Normal bolus during a square wave bolusUse this feature if you need additional insulin. You can program a normal bolus during the deliver

Página 124 - Cleaning your remote control

Bolus Menu 45 4. The bolus amount will flash. Use the and buttons to set the bolus. Press ACT. 5. The BOLUS DELIVERY screen will appear and will

Página 125 - Chapter 9

46 Chapter 4 Dual wave bolusA dual wave bolus is a combination of a normal bolus and a square wave bolus. Programming a dual wave bolus will deliver a

Página 126 - 114 Chapter 9

Bolus Menu 47 3. Select Dual Wave Bolus. Press ACT.4. The SET DUAL BOLUS screen will appear. The bolus amount will flash. Use the and buttons to s

Página 127 - I submerged my pump in water

ii Blood glucose targets . . . . . . . . . . . . . . . . . . . . . . . . 16Chapter 2Getting Started . . . . . . . . . . . . . . . . . . . . . . .

Página 128 - My screen appears distorted

48 Chapter 4 6. The duration icon will flash. Use the button to set the length of the delivery, in 30 minute increments. Press ACT.7. The BOLUS DEL

Página 129 - Pump maintenance

Bolus Menu 49 Delivering an express bolusThe EXPRESS BOLUS button is a short-cut to the Bolus Type menu.When square/dual wave bolus option is off1.

Página 130 - 118 Chapter 9

50 Chapter 4 When square/dual wave bolus option is on1. From the HOME screen, press the button. 2. The Bolus Type menu will appear.3. Select the

Página 131 - Storing your pump

Bolus Menu 51 2. Select Max Bolus. Press ACT.3. The current maximum bolus will flash. Use the and buttons to set the maximum amount and press ACT.

Página 132 - Special circumstances

52 Chapter 4 Easy bolusWhen using vibrate mode, Easy Bolus is limited to 20 steps or maximum bolus, whichever comes first. An Easy Bolus is an optiona

Página 133

Bolus Menu 53 programmed. Each beep is a different tone. This makes it easy for you to count the beeps, which equal the number of steps you have progr

Página 134 - 122 Chapter 9

54 Chapter 4 4. On/Set will be selected. Press ACT. If you do not wish to set the Easy Bolus Option, use the button and select OFF. Press ACT. If y

Página 135 - User Safety

Bolus Menu 55 Delivering an easy bolusThe feature is used to easily deliver a Normal Bolus. To deliver other types of boluses, use the Main menu or th

Página 136 - Warnings

56 Chapter 4 Pressing the or ESC button will cancel the programming. 3. Press ACT when the desired bolus amount is reached. The pump will beep or v

Página 137 - General precautions

Bolus Menu 57 1. From the HOME screen, press ACT. Bolus will be selected. Press ACT. 2. Select Bolus History. Press ACT.3. The BOLUS HISTORY screen

Página 138 - 126 Chapter 10

iii Easy bolus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52Setting up easy bolus . . . . . . . . . . . . . . . . . . .

Página 140 - Precautions - hyperglycemia

59 Chapter 5• • • • • • Basal Basal rate is the amount of insulin that the body needs to main-tain target glucose values without food.The basal ra

Página 141 - User Safety 129

60 Chapter 5 Basal menuThe menu choices that appear on the BASAL MENU screens will change if a temporary basal rate has been set or if the Patterns Op

Página 142 - Adverse reactions

Basal 61 What are start and stop times?Start and stop times are the times that one basal rate stops and the next basal rate starts. It is recommended

Página 143 - Site infection/abscesses

62 Chapter 5 Delivering a basalIt is recommended that you set basal rates with the assistance of your healthcare professional.1. From the HOME screen

Página 144 - 132 Chapter 10

Basal 63 4. The screen will change to Basal Rate 2. The start time for rate 2 will flash. If you require only one basal rate, press ACT again.5. The

Página 145 - User Safety 133

64 Chapter 5 Maximum basal rate (max basal) A maximum basal rate is a safety limit for the amount of basal insulin that is able to be delivered per ho

Página 146 - 134 Chapter 10

Basal 65 2. Select Max Basal Rate. Press ACT.3. The maximum basal rate will flash. Use the and buttons to change the rate. Rate range is 0.0 to 3

Página 147 - Alarms and Alerts

66 Chapter 5 How does it work? While it is active, a temporary basal rate overrides all other basal rate programming. When the temporary basal deliver

Página 148

Basal 67 4. The temporary basal rate will flash. Use the and but-tons to set the desired basal rate. Press ACT.Verifying temporary basal delivery

Página 149

iv Chapter 7Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 150 - ESC,ACT to clear

68 Chapter 5 Canceling a temporary basal rateCancels a temporary basal rate and returns to the programmed basal rate delivery.1. From the HOME screen

Página 151

Basal 69 Personal basal patternsThe personal basal patterns feature is optional for pump users. You may want to explore this option after you become f

Página 152

70 Chapter 5 Turning on personal basal patternsYour pump is programmed at the factory with this feature turned OFF. Discuss what uses and settings are

Página 153 - Use 1 AAA alkaline

Basal 71 Delivering personal basal patternsPersonal Basal Patterns allow you to set different basal delivery patterns. 1. From the HOME screen, pres

Página 154 - Message in Alarm History

72 Chapter 5 4. The SET BASAL RATE screen will appear. Set the Basal Rates desired, then press ACT. 5. The BASAL RATE screen will appear. The screen

Página 155 - Pump Specifications

Basal 73 Selecting a patternThis feature allows delivery of a personal basal pattern. Only pat-terns that have been programmed can be selected.1. Fro

Página 156 - 144 Chapter 12

74 Chapter 5 Basal reviewBasal Review provides a review of daily basal insulin deliveries. You should review your basal rates to make sure they match

Página 157 - Normal bolus

Basal 75 4. If you do have Patterns set, you will see the screen below. The current setting will be highlighted. Use the and but-tons to select t

Página 159 - Pump Specifications 147

77 Chapter 6• • • • • • SuspendThe Suspend function stops all insulin deliveries (Basal, Bolus, and Fixed Prime) that you have programmed into you

Página 160 - Water tight

v Chapter 9Troubleshooting & Care . . . . . . . . . . . . . . . . . 113Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 161 - Factory default settings

78 Chapter 6 Suspending the pumpThe pump will beep/vibrate in fifteen minute intervals starting at the closest quarter hour mark while in the Suspend

Página 162 - Icon table

Suspend 79 4. After 30 seconds, your pump will return to the HOME screen which now will show solid circles, indicating your pump is in the Attention

Página 164 - 152 Chapter 12

81 Chapter 7• • • • • • UtilitiesThe Paradigm pump can be programmed from the Utilities Menu to perform different functions for your safety and co

Página 165 - Index

82 Chapter 7 You can review your past alarms in the alarm history screen. Alarms result in closed circles on the screen. Pressing any button will re

Página 166

Utilities 83 2. The Alarm option will be selected. Press ACT.3. History will be selected. Press the button to select Alert Type. Press ACT.4. Sel

Página 167

84 Chapter 7 Using Auto-offProvides a safety mechanism that stops insulin delivery after a defined period of time (1 to 24 hours) if no button presses

Página 168

Utilities 85 insulin, and an alarm will sound after 12 hours if no buttons are pressed. Press ACT.You can set Auto Off from 0 (off) to 24 hours.Stop a

Página 169

86 Chapter 7 2. The Alarm option will be selected. Press ACT.3. The History option will be selected. Press ACT.4. Use the and buttons to review

Página 170

Utilities 87 Reviewing the daily totalsAllows you to review the total amount of insulin that you have received in each of the past 14 days. The daily

Comentários a estes Manuais